Researchers Compare Standard Cleaning to Passive Disinfection Cap for Decontamination of Needle-Free Connectors
The optimal decontamination method for needle-free connectors is still unresolved. The objective of this study by
The injection ports of needle-free connectors were inoculated with Staphylococcus aureus and allowed to dry. Disinfection caps containing 70% (v/v) isopropyl alcohol (IPA) were attached to the connectors for one, three or seven days and were compared with needle-free connectors cleaned with 2% (w/v) chlorhexidine gluconate (CHG) in 70% (v/v) IPA. The number of S. aureus remaining on the injection ports was evaluated. Median log10 reductions and 95% confidence interval (CI) were calculated and data analyzed using the Mann-Whitney test.
The application of the disinfection cap resulted in a significantly higher reduction in S. aureus than the 2% (w/v) CHG in 70% (v/v) IPA wipe, achieving a > 5 Log10 reduction in CFU at each time point.
The disinfection caps resulted in a significantly higher reduction in S.aureus on the injection ports when compared to the use of a 2% (w/v) CHG in 70% (v/v) IPA wipe. This offers an explanation for the lower rates of central-line associated bloodstream infection (CLABSI) associated with the use of disinfection caps reported in clinical studies.
Reference: Casey AL, Karpanen TJ, Nightingale P and Elliott SJ. An in vitro comparison of standard cleaning to a continuous passive disinfection cap for the decontamination of needle-free connectors. Antimicrobial Resistance & Infection Control.2018;7:50
Newsletter
Related Articles
- Bug of the Month: I'm Older Than Empires
September 16th 2025
- Top 5 Infection Prevention Articles of Summer 2025
September 16th 2025
- From Outbreak to Zoopocalypse: 11 More Must-Watch Viral Thrillers
September 15th 2025
- Debunking the Mistruths and Misinformation About COVID-19
September 15th 2025
- Bug of the Month: I Like to Hitch a Ride
September 12th 2025