Despite Safety Alerts, Contaminated Flexible Endoscopes Continue to Put Patients at Risk
From 2014 to 2021, continued FDA reports have shown multidrug-resistant bacteria contaminating 6 types of endoscopes have had fatal consequences.
Reprocessing practices may not always be sufficient to prevent flexible endoscopes from exposing patients to infectious organisms, including
Although health care workers reportedly cleaned and disinfected the endoscopes in compliance with manufacturer instructions, subsequent deaths were tied back to infection with CRE. Reprocessing is defined as pre-cleaning, leak-testing, and brushing the endoscope, followed by high-level disinfection or sterilization, and proper water-rinsing, drying, and storage.
“My analysis’ findings suggest that additional vigilance and enhanced reprocessing measures are integral to patient safety and preventing different types of flexible endoscopes from remaining contaminated and transmitting diseases,” investigator and author Lawrence F. Muscarella, PhD, of LFM Healthcare Solutions, LLC, told Infection Control Today® in an exclusive interview. “With the emergence of organisms resistant to multiple classes of antibiotics and antimicrobials—what we call today “superbugs”—along with an aging population who can benefit from treatments that involve use of a flexible endoscope but whose immune systems are less active, the game has changed, the runway shortened, and the risk of endoscope transmitting diseases is no longer academic but at times, fortunately still with low frequency, can be a matter of life and death.”
For the study, Muscarella analyzed 10,000 to 15,000 adverse event reports filed between 2014 and 2021 in the FDA’s
“Despite
According to the reports analyzed, contamination was identified on 6 different types of reprocessed, “patient-ready” endoscopes during microbiological surveillance testing. These included GI endoscopes such as colonoscopes, duodenoscopes, and gastroscopes; bronchoscopes; ear-nose-throat (ENT) endoscopes, including tracheal intubation endoscopes; and urological endoscopes, including cystoscopes.
Muscarella’s results suggest that unless specific additional safety measures are adopted and executed, such as inspecting the device for contamination and for damage before reuse, among several other sometimes overlooked safety measures, current reprocessing practices may not always be sufficient to prevent a flexible endoscope from transmitting infections. However, infection preventionists face challenges and the cleanings should be tracked, as ICT® reported
Relevant adverse events satisfying the requirements of the study included reports describing contamination of the reprocessed endoscope with microorganisms, patient fluids and/or biofilms, the documented association between infection and patient exposure to a flexible endoscope, failure to clean or otherwise reprocess the endoscope correctly according to manufacturer instructions, and the use of a damaged, cracked or otherwise broken endoscope that could pose an infection risk.
Key study results that surprised Muscarella the most were “that the number of adverse event reports submitted to FDA last year [that], satisfying [my] analysis’ inclusion criteria, compared [with] the previous year (2020), increased for each of the 6 studied endoscope types—and most markedly for ENT endoscopes,” the investigator told ICT®. Finally, “the number of adverse event reports submitted to the FDA in 2021 satisfying [my] study’s inclusion criteria, compared with 7 years earlier in 2014, increased significantly for each of these 6 endoscope types, and with gastroscopes displaying the most pronounced increase.”
Articles in this issue
over 3 years ago
Article
The FDA Issues Urgent Warning About Karl Storz Endoscopes Reprocessing MethodsNewsletter
Related Articles
- Bug of the Month: I'm Older Than Empires
September 16th 2025
- Top 5 Infection Prevention Articles of Summer 2025
September 16th 2025
- From Outbreak to Zoopocalypse: 11 More Must-Watch Viral Thrillers
September 15th 2025
- Debunking the Mistruths and Misinformation About COVID-19
September 15th 2025
- Bug of the Month: I Like to Hitch a Ride
September 12th 2025